HIOKI CT6875A - AC/DC Current sensor, 500 A / 2 MHz
Product information HIOKI CT6875A - AC/DC Current sensor, 500 A / 2 MHz
The HIOKI CT6875 AC/DC Current Sensor is designed for accurate current measurements across a broad frequency range, from DC to 2 MHz (Model CT6875-01: DC to 1.5 MHz). With a rated current of 500 A, it is ideal for R&D, quality evaluation, manufacturing, and maintenance in industries such as wireless charging systems, inverter motors, PV power conditioners, and EV quick charging facilities. In combination with the Power Analyzer PW8001, the sensor allows for precise high-frequency and low power factor measurements.
- Rated Current: 500 A
- Frequency Bandwidth: CT6875: DC to 2 MHz
- Basic Accuracy: Amplitude: ±0.04 % rdg. ±0.008 % f.s., Phase: ±0.08°
- Maximum Conductor Diameter: 36 mm
- Maximum Rated Voltage to Ground: 1000 V CAT III
- Operating Temperature: -40°C to 85°C
- Input Impedance Requirement: 1 MΩ or higher
- Output Terminal: ME15W
- Cable length: 3m
More than specs – find more details here:
Measure Power with Precision – Across All Applications
| Model | Rated current | Bandwidth | Basic accuracy | Max. Conductor diameter |
|---|---|---|---|---|
| CT6877A | 2000 A | DC – 1 MHz | 0,04 % rdg | Ø 80 mm |
| CT6876A | 1000 A | DC – 1.5 MHz | 0,04 % rdg | Ø 36 mm |
| CT6875A | 500 A | DC – 2 MHz | 0,04 % rdg | Ø 36 mm |
| CT6904A | 500 A | DC – 4 MHz | ±0,025 % rdg | Ø 32 mm |
| CT6873 | 200 A | DC – 10 MHz | ±0,03 % rdg | Ø 24 mm |
| CT6872 | 50 A | DC – 10 MHz | ±0,03 % rdg | Ø 24 mm |
| CT6863-05 | 200 A | DC – 500 kHz | ±0,05 % rdg | Ø 24 mm |
| CT6862-05 | 50 A | DC – 1 MHz | ±0,05 % rdg | Ø 24 mm |
Dimensions & Packaging
| Width | 500 mm |
|---|---|
| Height | 1,120 mm |
| Weight | 1.06 kg |