Measuring Ultrafast Inverters with High Precision
The advancement of inverters is a key factor in expanding electromobility and transitioning to renewable energies. Much depends on their ability to convert DC power from batteries and PV systems into usable AC power. Switching speed is an important metric. Another is efficiency. Specifically, in analyses of ultrafast SiC or GaN-based inverters, a peculiar phenomenon some-times occurs: it seems that more AC power is output than DC power is input. Such a physical paradox immediately stands out as an error. It is more challenging when a measurement shows a realistic but erroneous value. The goal is to avoid measurement errors like these. This is only possible with measuring instruments that meet the demands of new technologies. These are power measurement systems with reliable phase error correction.

The reliability of many power analysers ends with SiC and GaN technologies
Phase errors have hardly any effect up to switching frequencies of around 10 kHz. Beyond this limit, however, many power analysers calculate incorrect efficiencies. With SiC-based inverters, tenfold switching frequencies of up to 100 kHz are now possible. GaN-based inverters even switch with a factor of 100 up to 1 MHz. As a result, commercially available power analysers produce unreliable measurements of the phase difference. They also use current sensors from third-party suppliers. The HIOKI solution always combines a HIOKI power analyzer with exter-nal HIOKI zero flux current sensors, which are designed for accurate power measurement and that have unique features to optimize high frequency power measurement. The most important feature the phase shift compensation function to reduce the influence of phase errors.

Automatic Phase Shift Correction Minimizes Phase Errors
To correct measurement errors in the phase difference at high switching frequencies, HIOKI has developed a particularly effective phase shift correction for the PW8001 power analyser. For a phase shift compensation like that to work properly two things are necessary:
• A power analyzer that correctly performs phase correction
• A zero-flux current sensor with known time delay

The new phase shift correction is comparable to the familiar deskew function in oscilloscopes: If two different signals arrive at the oscilloscope at different times due to latencies then the “deskew” function lets you align those signals by compensating the latency with a fixed time val-ue. The phase shift compensation is basically the same thing because the phase error is directly re-lated to the time delay of the current sensor. For a current sensor of the CT68xxA series from HI-OKI, for example, the delay is shown in Figure 3. You can see the time delay in nanoseconds de-pending on the frequency.

It is important to know: The 100ns delay at 100Hz doesn’t have the same impact as 100ns de-lay at 1MHz. This becomes clear when transforming the above time delay into phase delay values described in degrees:
Analysers and sensors work together precisely
To address this phase delay, HIOKI has developed its zero-flux current sensors together with its Power Analyzers. To efficiently compensate for phase delay, the time delay of the current sensor must remain constant regardless of the frequency. This is the case with the HIOKI current sen-sors that also communicate the value of the phase delay automatically to the power analyzer. In addition, the most important key figures of the current sensors are automatically transmitted to the Power Analyser via Plug & Play. That includes the sensor type, the serial number and the measuring range. The current sensor only needs to be connected to the PW8001 Power Ana-lyser in order to be used without errors and to automatically compensate for phase errors. Re-turning to the "deskew" function: Simply connecting the current sensor to the PW8001 power analyzer allows the phase error to be automatically compensated.
The following behaviour can be observed with commercially available current sensors:

With current sensors from other manufacturers, the time delay varies depending on the frequen-cy. Which value can be used as the “deskew” parameter? Phase shift correction is virtually im-possible with non-system sensors.
No room for error
Another feature that makes HIOKI current sensors unique is that the position of the wire core does not influence the phase delay:

The reason why you can only see one single line in the chart Figure 6 is because the phase de-lay curves for all five measurement positions are the same, which does not leave room for er-rors. Not all current sensors offer this feature as standard. As shown in diagram 7 below, the conductor position in the sensor has an effect on the phase error in third-party sensors.

High-voltage dividers are the optimum solution for applications up to 5 kV
If you are not only working with high frequency but also higher voltage applications, there is a solution that further covers this area: a highly accurate high voltage divider. With the capability to precisely measure power up to 5 kV, it represents a significant advancement in accuracy com-pared to existing solutions utilizing differential probes. This is particularly beneficial for high volt-age and high frequency applications involving SiC and GaN semiconductors because of its 4MHz bandwidth and a known phase error which makes it possible to compensate for the phase error in the power analyzer.
Reproducible measurement results
Repeatability of measurement results is essential to draw conclusions. HIOKI Engineers at our HQ in Japan have measured the efficiency of a SiC based inverter using the PW8001 Power Analyzer in combination with the 50A AC/DC current sensors CT6872 over a period of 30 sec-onds. The switching frequency of the SiC semiconductors was 50kHz, and the motor was at a constant speed of 1000rpm. The integration time for the measurement was set to 200ms and the motor current in this experiment was approximately 2A. The graph below shows the excel-lent stability of the measurement:

The difference between the maximum and minimum result during the 30 seconds was less than 0.2 %. This experiment demonstrates that even at low currents this setup delivers excellent re-peatability, which is key in the development stage because it enables to identify even small effi-ciency gains due to changes in the design. However, such accurate results can only be achieved with an effective phase shift compensation. Under the same test conditions two well-known suppliers of power measurement systems have been tested and the outcome is shown in figure 9:

In this case, the repeatability of the measurements is a lot worse. For both systems the differ-ence between the maximum and minimum efficiency is more than 1%. This makes it very hard for engineers to conclude if the changes they make in their designs have an effect or not. For instance, even if you measure an increase in efficiency of 0.9%, this might simply be caused by the poor measurement repeatability, and you might not have achieved an increase in efficiency after all.
A measurement system from one source
Effective phase shift compensation requires the combination of a power analyser with a phase shift compensation function and a current sensor with stable and known time delay characteris-tics at high frequencies. The PW8001 Power Analyzer even features automated phase shift compensation, allowing it to automatically recognize the current sensor and adjust the phase compensation accordingly using the data provided by the sensor.

HIOKI has been designing and producing current sensors for power measurement for many years and time delay characteristics have always been a design focus for the engineers. On the other hand, zero flux current sensors from other manufacturers are typically designed for accu-rate current sensing where time delay characteristics are less important. Therefore, only HIOKI power analysers in conjunction with HIOKI zero-flux current sensors measure high-frequency inverters with absolute precision. To measure efficiency accurately, both the DC input power and the 3-phase output power must be measured precisely.